Test Code HAVM or LAB798 Hepatitis A IgM, Serum
Specimen Requirements
Container Type: |
Optimal: Gold Acceptable: Red.
|
Optimal Collection Volume: |
4.5mL; full tube |
Minimum Volume: |
1 mL whole blood |
Collection Instructions: |
Gently invert the sample 5-6 times after collection. |
Specimen Transport: |
Refrigerate |
Processing Instructions: |
1. Allow the specimen to clot for 30 minutes. 2. Centrifuge specimen within 2 hours of collection. 3. After centrifugation, specimen tubes without a gel barrier should have the serum aliquoted to a false bottom container. 4. Keep the serum refrigerated until testing can be performed. |
Specimen Stability
Specimen Type | Temperature | Time |
---|---|---|
Serum | Refrigerated | 7 days |
Frozen | 3 months |
Specimen Rejection Criteria
Hemolysis | Mild OK; Moderate OK; Gross Reject |
Lipemia | Mild OK; Moderate OK; Gross Reject |
Icterus | Mild OK; Moderate OK; Gross Reject |
Other | Quantity not sufficient |
Useful For
This assay is intended for use as an aid in the diagnosis of an acute or recently acquired hepatitis A virus infection.
Methodology
Roche Cobas - Electrochemiluminescence immunoassay
Reference Ranges
Negative
Day(s) Performed
24/7
Expected TAT
Same day
Performing Laboratory
Bronson Laboratory, Chemistry - Kalamazoo
Sample Retention Time
7 days
CPT Code
86709
LOINC Code
22314-9