Test Code CBGA or LAB2113 Blood Gas Cord, Arterial
Profile Information
Test ID |
Reporting Name |
Available Separately |
Always Performed |
---|---|---|---|
TEMP* |
Temperature |
N | Y |
FIO2* |
Inspired O2 |
N | Y |
HGB |
Hemoglobin |
N |
Y |
PH, Cord Art |
PH, Cord Arterial |
N | Y |
PCO2, Cord Art |
pCO2, Cord Arterial |
N | Y |
PO2, Cord Art |
pO2, Cord Arterial |
N | Y |
HCO3 |
HCO3 |
N | Y |
OXYHGB, Cord Art |
Oxyhemoglobin, Cord Arterial |
N |
Y |
BBE or BBD |
Base Excess or Base Deficient |
N |
Y |
*Captured during specimen collection.
Specimen Requirements
Container Type: |
Heparinzed syringe, whole blood
|
Optimal Collection Volume: |
1 syringe (2 mL whole blood) |
Minimum Volume: |
0.5 mL |
Collection Instructions: |
|
Specimen Transport: |
Refrigerate |
Processing Instructions: |
|
Specimen Stability
Assay as soon as possible. Specimens are stable at room temperature for 30 minutes and 2 hours on ice.
Specimen Rejection Criteria
Quantity not sufficient
Methodology
Radiometer ABL 825
- pH and pCO2 - Potentiometric
- pO2 - Amperometric
- HCO3 - Calculated result
- Hemoglobin parameters - Optical system
Reference Ranges
Test ID |
Reference Range and Reporting Units |
---|---|
HGB |
Gender and age-specific* |
PH, Cord Art |
7.13-7.41 |
PCO2, Cord Art |
28-73 mm[Hg] |
PO2, Cord Art |
15-27 mm[Hg] |
HCO3 |
14-28 mmol/L |
OXYHGB |
18-51 % |
Please click the Bronson Critical Values link to view Critical Values
* For HGB reference ranges, please link the Hematology Reference Ranges link.
Day(s) Performed
24/7
Expected TAT
Same day
Performing Laboratory
Bronson Laboratory, Chemistry - Kalamazoo
Bronson Respiratory - Battle Creek
Sample Retention Time
Samples are discarded once per shift (every 8 hours)
CPT Code
82805