Sign in →

Test Code UUNR or LAB748 Urea Nitrogen (BUN), Random, Urine

Specimen Requirements

Container Type:

Sterile urine cup

 

 

Optimal Collection Volume: 

5 mL

Minimum Volume:

1 mL

Collection Instructions:

1. Collect a random urine specimen from the patient.

2. Ensure the lid is secured tightly before sending it to the lab.

3. Ensure that no preservatives are used.

Specimen Transport:

Refrigerate

Processing Instructions: 

1. Pour off urine into a labeled aliquot tube.

2. Centrifuge the urine aliquot.

3. Deliver the urine aliquot to the chemistry department.

Specimen Stability

Specimen Type Temperature Time
Urine Refrigerated 7 days
Room Temperature 2 days
Frozen 1 month

Specimen Rejection Criteria

Quantity not sufficient

Useful For

Assessment of kidney failure (prerenal vs. acute kidney injury)

Methodology

Roche Cobas - Kinetic test with urease and glutamate dehydrogenase

Reference Ranges

No established reference values.

Day(s) Performed

24/7

Expected TAT

Same day

Performing Laboratory

Bronson Laboratory, Chemistry - Kalamazoo and Battle Creek

Sample Retention Time

7 days

CPT Code

84540

LOINC Code

3095-7