Sign in →

Test Code GLDG or LAB2177 Gliadin Deamidated IgG, Serum

Specimen Requirements

Container Type:

Optimal: Gold

Acceptable: Red

Optimal Collection Volume: 

4.5 mL; full tube

Minimum Volume: 2 mL
Collection Instructions:

Gently invert the sample 5-6 times after collection.

Specimen Transport:

Refrigerate

Processing Instructions: 

1. Allow sample to clot for 30 minutes.

2. Centrifuge specimen within 2 hours of collection.

3. Separate serum into a false bottom container.

4. Keep serum refrigerated until testing can be performed.

Specimen Stability

Specimen Type Temperature Time
Serum Refrigerated 7 days
Room Temperature 8 hours

 

Specimen Rejection Criteria

Hemolysis Mild OK; Moderate OK; Gross Reject
Lipemia Mild OK; Moderate OK; Gross Reject
Icterus Mild OK; Moderate OK; Gross Reject
Other Quantity not sufficient

Useful For

Used for patients with an IgA deficiency, who can test false-negative with the IgA form of Gliadin Deamidated.

Methodology

Bio-Rad BioPlex 2200: Multiplex flow immunoassay

Reference Ranges

 

 Test ID   Reference Range and Reporting Units
GLDG < 15 U/mL: Negative
> 15 U/mL: Positive

 

Note: Reference range changed on 1/22/19. Please see the link to January 2019 LabWire under "Resources" on this page for details.

Day(s) Performed

Monday through Friday

Expected TAT

Same day or 1-2 days

Performing Laboratory

Bronson Laboratory, Chemistry - Kalamazoo

Sample Retention Time

7 days

CPT Code

83516

LOINC Code

63459-2